The Leybold SCROLLVAC 3S plus is a compact dry scroll pump for small laboratory systems and suitable instrument backing duties. It provides 3.0 m³/h at 50 Hz or 3.5 m³/h at 60 Hz. The scientific 3S version adds an inlet isolation valve and delayed opening to protect the connected system.
Leybold SCROLLVAC 3S plus illustrated with inlet isolation valve. The standard 3 plus has a different inlet arrangement.
The Leybold SCROLLVAC 3S plus is a compact dry scroll pump for small laboratory systems and suitable instrument backing duties. It provides 3.0 m³/h at 50 Hz or 3.5 m³/h at 60 Hz. The scientific 3S version adds an inlet isolation valve and delayed opening to protect the connected system.
Key specifications
Specification
Details
Peak pumping speed, 50 / 60 Hz
3.0 / 3.5 m³/h
Ultimate pressure
0.1 mbar gas ballast closed; 0.15 mbar one turn open
Electrical supply
Voltage-specific single-phase 100, 115, 200 or 230 V; 50/60 Hz
Inlet / exhaust
DN16 ISO-KF / 1/8-inch BSP, 6 × 8 mm hose
Dimensions, L × W × H
224 × 158 × 231 mm
Maximum mass
8.0 kg
Ambient operation
5–40 °C, indoor
Power consumption at ultimate
210 W
Inlet isolation
Factory-fitted inlet valve and delay relay
Equipment condition
New; configured by quotation
Ultimate pressure is a manufacturer test value, not a guaranteed pressure at the chamber. Pipe conductance, gas load, temperature and condition affect installed performance.
Documented configurations
Specification
Details
142030V430
230 V single-phase, 50/60 Hz
142030V420
200 V single-phase, 50/60 Hz
142030V410
115 V single-phase, 50/60 Hz
142030V400
100 V single-phase, 50/60 Hz
Standard 3 plus or scientific 3S plus?
Both versions have a built-in exhaust valve. The 3S adds a fast-acting solenoid inlet valve and a relay that delays opening by 10 seconds after power is restored. The standard 3 plus has no built-in inlet valve. The manufacturer brochure lists an exhaust filter as supplied with 3S; confirm the complete supplied-equipment list in the quotation.
Power and installation
Request the 230 V version for UK single-phase installations. The voltage choices are separate pump builds, not a universal-voltage input. Specify the separately selected UK lead 800102V0003. Keep ventilation clear, support the pipework and follow the permitted exhaust back-pressure limits.
Gas ballast, accessories and maintenance
One turn of gas ballast gives 2.5 l/min flow and a 0.15 mbar ultimate-pressure rating. Use inlet protection and a compatible exhaust arrangement for the process. Silencer/filter accessories and service kits are selected separately unless expressly included. The pump is dry in its vacuum space, but bearings contain isolated lubricant and tip seals require maintenance. Prolonged operation at high inlet pressure can reduce bearing and tip-seal life.
The 3S inlet valve contains brass in the gas path. State the gas and chemical composition when enquiring so material compatibility can be checked.
Dimensions and installation
leybold scrollvac 3 dimensions; original manufacturer drawing page, with units and configuration labels retained. Allow the specified ventilation and service clearances. Select to enlarge.
Source: SCROLLVAC 3/3S brochure 3610 0310 02, pages 6–12; operating instructions 301098335_002_C3, pages 13, 16–21. This model-level listing covers the documented choices; the exact order reference, included equipment, availability and lead time are confirmed with the quotation.
For SCROLLVAC 3S plus, include working pressure, required flow or chamber volume, gas and vapour load, operating hours, electrical supply and connections. For a replacement, send the complete nameplate and serial number so the configuration can be checked.
Confirm the exact version, electrical supply and connections. Ask whether oil, filters, fittings, cables or controllers are included or separately specified.
Availability, lead time and the applicable warranty are confirmed for your quotation. A catalogue listing does not reserve stock.